Y560-A Self-cleaning Ammonium Sensor Calibration


Y560-A Self-cleaning Ammonia Sensor Calibration

Required equipments and raw materials
Required equipments: three 1000mL beakers, one set of gloves.
Raw materials: 10mg/L ammonium ion electrode standard solution, 100mg/L ammonium ion electrode standard solution and deionized water.

1701842995209914.png

Calibration steps
There are two methods of user calibration for the ammonium ion electrode of the ammonia nitrogen sensor: 1-point calibration and 2-points calibration.
1-point calibration
was done in 10mg/L standard solution, 2-points calibration was done in 10mg/L standard solution and 100mg/L standard solution.

1-point calibration step (change the K value)

1701843400139586.png

Step 1: put an appropriate amount of 10mg/L ammonium ion electrode standard solution into a 250mL beaker, add an ionic regulator solution in the corresponding proportion (2% of the total amount of solution).
Step 2: connect the cable to the sensor, remove the sensor protective cover, remove the protective sleeves of the ammonium ion electrode, pH electrode and reference electrode, and then install the sensor protective cover. Put it into the  beaker containing 10mg/L ammonium ion electrode standard solution, and insert the RS485 to USB connector connected to the sensor cable into the computer.
Step 3: open the SmartPC software on the computer, then click “Calibrate”, and click “Get” under the “Manual Calibration” interface to get the initial K and B values of the sensor. Under the “Automatic Calibration” interface, change the “Single/First Point Calibration Value” to 10, click “1 points” to do one-point calibration, wait for the values in the pop-up window to stabilize, and click “OK” to complete the calibration. Click “Get” to check whether the K value is written correctly.
Step 4: click “Start” under the “NH4-N” interface to get the real-time measurement of the pH、NH4+、NH4-N and temperature, check that the NH4+ is close to 10mg/L to verify that the one-point calibration is successful.

2-points calibration step (change the K and B values)

1701843708366487.png

Step 1: put an appropriate amount of 10mg/L ammonium ion electrode standard solution into beaker, add an ionic regulator solution in the corresponding proportion (2% of the total amount of solution). Put an appropriate amount of 100mg/L ammonium ion electrode standard solution into beaker, add an ionic regulator solution in the corresponding proportion (2% of the total amount of solution).
Step 2: connect the cable to the sensor, remove the sensor protective cover, remove the protective sleeves of the ammonium ion electrode, pH electrode and reference electrode, and then install the sensor protective cover. Put it into the  beaker containing 10mg/L ammonium ion electrode standard solution, and insert the RS485 to USB connector connected to the sensor cable into the computer.
Step 3: open the SmartPC software on the computer. Under the “Automatic Calibration” interface, change the “Single/First Point Calibration Value” to 10, change the “Second Point Calibration Value” to 100, click “2 points” to do two-points calibration, wait for the values in the pop-up window to stabilize, and click “OK”. Then clean the sensor with distilled water or deionized water, wipe it dry and put it into the beaker containing 100mg/L ammonium ion electrode standard solution, wait for the values in the pop-up window to stabilize, and click “OK” to complete the calibration. Click “Get” to check whether the K and B values are written correctly.
Step 4: click “Start” under the “NH4-N” interface to get the real-time measurement of the pH、NH4+、NH4-N and temperature, check that the NH4+ is close to 100mg/L, then clean the sensor with distilled water or deionized water, wipe it dry and put it into the beaker containing 10mg/L ammonium ion electrode standard solution, check that the NH4+ is close to 10mg/L to verify that the two-points calibration is successful.

Y561-A NH4-N ammonium sensor calibration

Y561-A NH4-N ammonia sensor calibration

Required equipments and raw materials
Required equipments: three 1000mL beakers, one set of gloves.
Raw materials: 10mg/L ammonium ion electrode standard solution, 100mg/L ammonium ion electrode standard solution and deionized water.

1701844759451699.jpg

Calibration steps
There are two methods of user calibration for the ammonium ion electrode of the ammonia nitrogen sensor: 1-point calibration and 2-points calibration.
1-point calibration
was done in 10mg/L standard solution, 2-points calibration was done in 10mg/L standard solution and 100mg/L standard solution.

1-point calibration step (change the K value)

1701844838152026.jpg

Step 1: put an appropriate amount of 10mg/L ammonium ion electrode standard solution into a 250mL beaker, add an ionic regulator solution in the corresponding proportion (2% of the total amount of solution).
Step 2: connect the cable to the sensor, remove the sensor protective cover, remove the protective sleeves of the ammonium ion electrode, pH electrode and reference electrode, and then install the sensor protective cover. Put it into the  beaker containing 10mg/L ammonium ion electrode standard solution, and insert the RS485 to USB connector connected to the sensor cable into the computer.
Step 3: open the SmartPC software on the computer, then click “Calibrate”, and click “Get” under the “Manual Calibration” interface to get the initial K and B values of the sensor. Under the “Automatic Calibration” interface, change the “Single/First Point Calibration Value” to 10, click “1 points” to do one-point calibration, wait for the values in the pop-up window to stabilize, and click “OK” to complete the calibration. Click “Get” to check whether the K value is written correctly.
Step 4: click “Start” under the “NH4-N” interface to get the real-time measurement of the pH、NH4+、NH4-N and temperature, check that the NH4+ is close to 10mg/L to verify that the one-point calibration is successful.

2-points calibration step (change the K and B values)

1701844965206804.jpg

Step 1: put an appropriate amount of 10mg/L ammonium ion electrode standard solution into beaker, add an ionic regulator solution in the corresponding proportion (2% of the total amount of solution). Put an appropriate amount of 100mg/L ammonium ion electrode standard solution into beaker, add an ionic regulator solution in the corresponding proportion (2% of the total amount of solution).
Step 2: connect the cable to the sensor, remove the sensor protective cover, remove the protective sleeves of the ammonium ion electrode, pH electrode and reference electrode, and then install the sensor protective cover. Put it into the  beaker containing 10mg/L ammonium ion electrode standard solution, and insert the RS485 to USB connector connected to the sensor cable into the computer.
Step 3: open the SmartPC software on the computer. Under the “Automatic Calibration” interface, change the “Single/First Point Calibration Value” to 10, change the “Second Point Calibration Value” to 100, click “2 points” to do two-points calibration, wait for the values in the pop-up window to stabilize, and click “OK”. Then clean the sensor with distilled water or deionized water, wipe it dry and put it into the beaker containing 100mg/L ammonium ion electrode standard solution, wait for the values in the pop-up window to stabilize, and click “OK” to complete the calibration. Click “Get” to check whether the K and B values are written correctly.
Step 4: click “Start” under the “NH4-N” interface to get the real-time measurement of the pH、NH4+、NH4-N and temperature, check that the NH4+ is close to 100mg/L, then clean the sensor with distilled water or deionized water, wipe it dry and put it into the beaker containing 10mg/L ammonium ion electrode standard solution, check that the NH4+ is close to 10mg/L to verify that the two-points calibration is successful.

Y514-A Self-cleaning Chlorophyll Sensor Calibration

Y514-A self-cleaning chlorophyll sensor calibration

Required equipments and raw materials
Required equipments: two 1L brown bottles, one iron stand, one set of gloves
Raw materials: 2mg/L chlorophyll standard solution and deionized water

1701845597153006.png

Calibration steps
There are three methods of user calibration for the chlorophyll sensor: 0-point calibration, 1-point calibration and 2-points calibration.
0-point calibration was done in deionized water, 1-point calibration was done in 2mg/L chlorophyll standard solution, 2-points calibration was done in deionized water and 2mg/L chlorophyll standard solution.

0-point calibration step (change the B value)

1701845786183662.jpg

Step 1: put an appropriate amount of distilled water or deionized water into a 1L brown bottle, remove the rubber protective cap on the front of the sensor and put the sensor in the center of the brown bottle, which not close to the wall around the brown bottle, and then fix the sensor with an iron stand, adjust the high and low position of the sensor. Pay attention to maintain a distance greater than 10cm between the front end of the sensor and the bottom of the brown bottle.
Step 2: Under the “Automatic Calibration” interface, click “0 point” to do zero-point calibration, wait for the values in the pop-up window to stabilize, and click “OK” to complete the calibration. Click “Get” to check whether the B value is written correctly.
Step 3: click “Start” under the “CHL” interface to get the real-time measurement of the CHL and temperature, check that the CHL is close to 0 to verify that the zero-point calibration is successful.

1-point calibration step (change the K value)

1701847883105333.jpg

Step 1: put an appropriate amount of 2mg/L chlorophyll standard solution into a 1L brown bottle, remove the rubber protective cap on the front of the sensor and put the sensor in the center of the brown bottle, which not close to the wall around the brown bottle, and then fix the sensor with an iron stand, adjust the high and low position of the sensor. Pay attention to maintain a distance greater than 10cm between the front end of the sensor and the bottom of the brown bottle.
Step 2: Under the “Automatic Calibration” interface, change the “Single/First Point Calibration Value” to 227.1, click “1 point” to do one-point calibration, wait for the values in the pop-up window to stabilize, and click “OK” to complete the calibration. Click “Get” to check whether the K value is written correctly.
Step 3: click “Start” under the “CHL” interface to get the real-time measurement of the CHL and temperature, check that the CHL is close to 227.1 to verify that the one-point calibration is successful.

2-points calibration step (change the K and B values)

1701848045131892.jpg

Step 1: put an appropriate amount of distilled water or deionized water into a 1L brown bottle, and put an appropriate amount of 2mg/L chlorophyll standard solution into another 1L brown bottle.
Step 2: remove the rubber protective cap on the front of the sensor, put the sensor in the center of the brown bottle containing distilled water or deionized water, which not close to the wall around the brown bottle, and then fix the sensor with an iron stand, adjust the high and low position of the sensor. Pay attention to maintain a distance greater than 10cm between the front end of the sensor and the bottom of the brown bottle.
Step 3: Under the “Automatic Calibration” interface, change the “Single/First Point Calibration Value” to 0, change the “Second Point Calibration Value” to 227.1, click “2 points” to do two-points calibration, wait for the values in the pop-up window to stabilize, and click “OK”. Then wipe the sensor dry and put it into 2mg/L chlorophyll standard solution according to the method of step 2, wait for the values in the pop-up window to stabilize, and click “OK” to complete the calibration. Click “Get” to check whether the K and B values are written correctly;
Step 4: click “Start” under the “CHL” interface to get the real-time measurement of the CHL and temperature, check that the CHL is close to 227.1, then clean the sensor with distilled water or deionized water, wipe it dry and put it into the distilled water or deionized water according to the method of step 2 to get the real-time measurement of the CHL and temperature, check that the CHL is close to 0 to verify that the two-points calibration is successful.